Practical Instrument Electronics
Model 520/521
T/C Simulator(s)
Operating Instructions
Basic Keypad Operations
1 EZ-Check™ Switch
Slide the switch to select from three user stored values
for the desired calibration points. The User can select HI,
DIAL and LO positions. These values can easily be
changed to suit the calibration requirements.
.
2 ON/OFF Switch
Slide the ON/OFF Switch to turn the Model
520/521 on or off.
3 EZ-Dial™ Knob
The EZ-Dial™ Knob has two adjustment
speeds. Simply turning the EZ-Dial™ Knob will
select fine adjustments. While pressing down
and turning the EZ-DialKnob will make
course adjustments.
Note: When the EZ-Check™ Switch is in the
HI or LO position, pressing and holding the EZ-
Dial™ Knob without turning
will store new HI
or LO values.
Model 520/521 Configuration
Instructions for Enabling and Disabling the Configuration Options
1. Turn the Model 520/521 on with the ON/OFF Switch2.
2. Press the EZ-Dial Knob
3 while the “PRESS EZ-DIAL KNOB FOR
CONFIGURATION” message is displayed.
3. Select options by turning the EZ-Dial Knob until the arrow points
to the desired option.
4. The option can be enabled or disabled by tapping the EZ-Dial
Knob.
The Model 520/521 configuration menu will exit automatically after 5
seconds of inactivity and go to normal operation with the options selected.
These options are recalled at turn on until they are changed again.
841 Holt Road #1 Webster, NY 14580 Tel: 585-872-9350 Fax: 585-872-2638 [email protected] http://www.piecal.com/
Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.
520/521-9002 2/25/04 1-4
Zobrazit stránku 0
1 2 3 4

Shrnutí obsahu

Strany 1 - T/C Simulator(s)

Practical Instrument Electronics Model 520/521 T/C Simulator(s) Operating Instructions Basic Keypad Operations 1 EZ-Check™ Switch Slide the switch

Strany 2 - Operating Instructions

Model 520/521 Operating Instructions Model 520/521 Configuration Model 520/521 Configuration Menu Auto Off ON (default)/OFF If Auto Off is ON,

Strany 3

Model 520/521 Operating Instructions Connection Diagram Two Wire Connection to Transmitter Specifications General Specifications: (Unless otherwi

Strany 4

Model 520/521 Operating Instructions 841 Holt Road #1 • Webster, NY 14580 Tel: 585-872-9350 • Fax: 585-872-2638 [email protected] • http://www.pi

Příbuzné modely 520

Komentáře k této Příručce

Žádné komentáře