Stepping and auto step/ramp limits are defined by the EZ-Check HI and LO settings. The step direction
changes when a limit is reached.
Figure 1 shows how the Step/Ramp Parameters are used to configure automatic stepping/ramping.
Figure 1
0
4
8
12
16
20
24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
time
mA
Note: The Model 532’s ability to detect overload/undervoltage conditions may be limited by the rate of
change in the output when using automatic stepping/ramping. Turn auto step/ramp off while connecting
or disconnecting the Model 532.
G. Quick Reference Bar Graph
The Quick Reference Bar Graph indicates the input and output level on the Model 532 in % of 4-20 mA
with 1% resolution.
step size = (EZ-Step HI – EZ-Step LO) ÷ # of steps
step time = ramp time ÷ # of steps