
Model 532 Operating Instructions
82 E. Main Street Suite 3.14 • Webster, NY 14580
Tel: 585-872-9350 • Fax: 585-872-2638
sales@piecal.com • http://www.piecal.com/
Practical Instrument Electronics, Inc. Copyright 2011. All rights reserved.
532-9002 Rev C 7/13/11
7-8
In all loop diagnostic modes, AC voltage is displayed. If more than 2 VAC is present on the Model 532’s
terminals, the display is highlighted to alert the user of a potential problem.
In all loop diagnostic modes, HART activity is indicated with “♥ HART DETECTED ♥” . The ♥ symbols
flash on and off with the actual HART messages.
J. Modes of Operation
Connect the Model 532 in series with
the process loop to monitor current.
Observe correct polarity. Current
limiting above 24.000 mA is indicated
by a flashing “CURRENT LIMITED”
display.
Power and Measure 2 Wire Transmitter
The Model 532 provides power to the process
loop while displaying output current. Use this
mode to test a transmitter’s ability to control
loop current. Current limiting above 24.000
mA is indicated by a flashing “CURRENT
LIMITED” display.
Read Volts
The Model 532 measures +/- 30 VDC with 4X
overrange ability. The display flashes
“OVERRANGE” when the 30 volt limit is
exceeded. 0.001 volt resolution below 24.00
volts provides the ability to calibrate 1-5 volt
instrumentation.
Komentáře k této Příručce